STABLE VACCINE TEMPERATURES WITH AUTO DEFROST FREEZER CONVENIENCE
Our proprietary “Controlled Auto Defrost” technology provides the convenience of an auto defrost design while still ensuring that vaccines and critical samples remain at consistent and stable temperatures below 5°F (-15°C)
ECO FRIENDLY NATRUAL REFRIGERANTS
The Controlled Auto Defrost Vaccine/Sample freezer uses hydrocarbon (natural) refrigerants that help reduce energy consumption, are non-toxic and have no ozone-depleting properties.
FEATURES AND BENEFITS
Click each feature (+) to reveal its benefits
Provides connectivity to central or independent alarm systems to notify responders who are off-site if an alarm condition occurs.
Environmentally-friendly refrigerant, no adverse effects on environment. Early compliance to US Federal Government Mandates. Uses less energy to remove heat from chamber, lowering energy expenses.
Removable and easily cleaned , design helps conduct cold and improves air circulation, improving temperature uniformity.
Longer warranty than industry standard, provides worry-free support through national network of field technicians.
Ability to incorporate independent locking and security systems
Meets hospital, clinic and pharmacy requirements for independent agency testing
Stronger, heavy-duty seal on door gasket keeps temperatures uniform, reduces energy consumption
Meets CDC guidelines to have an independent temperature monitoring device, reducing expense with a 3 year certificate of calibration, eliminating capital outlay for device recalibration.
Per CDC guidelines, every refrigerator or freezer should be equipped with a “Do Not Unplug” label as well as a warning label regarding changing controller settings.
Allows safe input of independent temperature monitoring device to measure chamber temperatures without compromising the door gasket.
CRITICAL TEMPERATURES ARE MAINTAINED, EVEN THROUGH MULTIPLE DAILY DEFROST CYCLES
With patented technology, the Lab Research Products Controlled Auto Defrost Freezer keeps samples and air temperature at or below the critical 5°F (-15°C) temperature range
Auto Defrost Cycle Every 6 Hours
At set point -11°F (-24°C)
Operating Temperature Range
5°F – -13°F (-15°C – -25°C)
Max Air Temperature
Empty Freezer: 5°F (-15°C)
Loaded Freezer: 3°F (-16°C)
Max Sample Temperature
Empty Freezer: 0°F (-18°C)
Loaded Freezer: -4°F (-20°C)