STABLE VACCINE TEMPERATURES WITH AUTO DEFROST FREEZER CONVENIENCE

Our proprietary “Controlled Auto Defrost” technology provides the convenience of an auto defrost design while still ensuring that vaccines and critical samples remain at consistent and stable temperatures below 5°F (-15°C)

Consistent Temperature Stability
CDC Compliant
Patented Frost-Free Technology
Environmentally Friendly Refrigerants

ECO FRIENDLY NATRUAL REFRIGERANTS

The Controlled Auto Defrost Vaccine/Sample freezer uses hydrocarbon (natural) refrigerants that help reduce energy consumption, are non-toxic and have no ozone-depleting properties.

FEATURES AND BENEFITS

Click each feature (+)  to reveal its benefits

No longer a need to relocate samples as would be required during a manual defrost process. No worry of product-damaging temperature excursions as would be experienced with an old technology/traditional auto defrost freezer

Provides connectivity to central or independent alarm systems to notify responders who are off-site if an alarm condition occurs.

Environmentally-friendly refrigerant, no adverse effects on environment. Early compliance to US Federal Government Mandates. Uses less energy to remove heat from chamber, lowering energy expenses.

Removable and easily cleaned , design helps conduct cold and improves air circulation, improving temperature uniformity.

Longer warranty than industry standard, provides worry-free support through national network of field technicians.

Ability to incorporate independent locking and security systems

Meets hospital, clinic and pharmacy requirements for independent agency testing

Stronger, heavy-duty seal on door gasket keeps temperatures uniform, reduces energy consumption

Meets CDC guidelines to have an independent temperature monitoring device, reducing expense with a 3 year certificate of calibration, eliminating capital outlay for device recalibration.

Per CDC guidelines, every refrigerator or freezer should be equipped with a “Do Not Unplug” label as well as a warning label regarding changing controller settings.

Allows safe input of independent temperature monitoring device to measure chamber temperatures without compromising the door gasket.

CRITICAL TEMPERATURES ARE MAINTAINED, EVEN THROUGH MULTIPLE DAILY DEFROST CYCLES

With patented technology, the Lab Research Products Controlled Auto Defrost Freezer keeps samples and air temperature at or below the critical 5°F  (-15°C) temperature range

Auto Defrost Cycle Every 6 Hours

At set point -11°F (-24°C)

Operating Temperature Range

5°F – -13°F (-15°C – -25°C)

Max Air Temperature

Empty Freezer: 5°F (-15°C)

Loaded Freezer: 3°F (-16°C)

Max Sample Temperature

Empty Freezer: 0°F (-18°C)

Loaded Freezer: -4°F (-20°C)